Specifications
Size: | Approx. 268 x 189 x 149 mm |
Weight: | Approx. 5400 g |
Base material: | Cellulose-polymer composite |
Optimal tube voltage: | 120 kVp (cf page 5) – adaptable upon request – |
This abdomen phantom can be used in CT for evaluation of low-contrast lesions in the liver. It was designed to enable evaluation of diagnostic software, including AI tools.
The phantom simulates a contrast medium enhanced abdomen in early portal venous phase and has 35 low-contrast liver lesions.
The phantom provides a detailed and realistic simulation of soft and bone tissue, including small details such as lymph nodes. Air voids are filled with a cellulose-polymer composite of approx. -80 HU.
The phantom can be used for detection, segmentation and classification tasks and other common methods of image quality evaluation.
Size: | Approx. 268 x 189 x 149 mm |
Weight: | Approx. 5400 g |
Base material: | Cellulose-polymer composite |
Optimal tube voltage: | 120 kVp (cf page 5) – adaptable upon request – |
35 rod-shaped liver lesions in 5 sections |
Lesion height: 10.9 mm |
Section 1: 8 lesions, 5 mm diameter, approx. -50 to -20 HU contrast at 120 kVp |
Section 2: 9 lesions, 13 mm diameter, approx. -40 to -20 HU contrast at 120 kVp |
Section 3: 9 lesions, 11 mm diameter, approx. -40 to -20 HU contrast at 120 kVp |
Section 4: 6 lesions, 8 mm diameter, approx. -30 to -20 HU contrast at 120 kVp |
Section 5: 3 lesions, 8 mm diameter, approx. -40 HU contrast at 120 kVp |